Abstract

In this study, we investigate the effects of pulsatile flow and inflow on dynamic susceptibility-contrast MRI intravascular arterial input function measurement in human brain arteries and measure how they are affected by first-order flow compensation. A dual-echo single-shot EPI sequence with alternating flow compensation gradients was used to acquire dynamic susceptibility-contrast images with electrocardiogram monitoring. The dynamic signal variations measured inside the middle cerebral and internal carotid arteries were associated to the pulsatile arterial blood velocities measured with a single-slice quantitative flow sequence throughout the cardiac cycle. Major inverse correlations between intravascular signal and blood velocity were found for the standard single-shot EPI sequence. Flow compensation reduces these correlated variations that contribute to signal physiological noise. This causes a significant twofold increase of intravascular SNR in the middle cerebral and the internal carotid arteries (2.3 ± 0.9, P = 0.03) and (2.0 ± 0.9, P = 0.04), respectively; and reduced phase SD for the internal carotid arteries (0.72 ± 0.14, P = 0.004). The correction proposed in this work translates into a quantitative arterial input function with reduced noise in the internal carotid arteries. The physiological noise added by pulsatile flow and inflow for intravascular arterial input function measurement in the brain arteries is significantly reduced by flow compensation.