Abstract

PurposeTo examine the influence of incorrectly placed gonad shields on radiation dose when performing abdominal CT with automatic exposure correction, using systems from different vendors. Methods and materialsAn anthropomorphic phantom was scanned without gonad shields, and with gonad shields placed in two different positions relative to the scan range. Dose Length Product was recorded. mA distribution in the longitudinal direction was plotted. Mean dose was compared using the t-test. ResultsThree scanners showed different increase in relative DLP according to shield position. ConclusionCare must be taken when placing lead shielding at CT and characteristics of each scanner should be known to the operator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.