Abstract

The purpose of this study was to asses the fate of incomplete seating of press-fit acetabular components. From 1990 to 2002, 587 hip arthroplasties were performed using a single-design acetabular component inserted with a press-fit technique with no additional fixation. Complete radiographic follow-up at a mean 9.2 years was available for 343 cases. One hundred thirty-three acetabular components had incomplete seating on the initial postoperative radiograph and 210 cases had complete seating. Of 133 components, 125 (94%) had evidence of complete zone 2 filling. Zone 2 focal osteolysis was present in 22 patients. There were no intraoperative fractures. No acetabular components were revised for loosening. Incomplete seating of press-fit acetabular components is safe and effective in achieving initial and long-term fixation. Zone 2 lucencies when present initially are not associated with increased failure risks. Vigorous attempts to completely seat the acetabular component appear to be unwarranted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.