Abstract

In PET, as the spatial resolution of the measurement system is increased, multiple interactions of a single photon may be separately measured and such events are often removed from the data used in image reconstruction. While for some PET imaging tasks this effect is unimportant, for primate, brain and high-spatial-resolution imaging where sensitivity is important, such Inter Crystal Scattering (ICS) events may constitute a large fraction of the measured data. On-the-fly list-mode image reconstruction is generally required in order to use all the information provided by ICS. Simulated One-Pass List-mode image reconstruction is used in this investigation to study the inclusion of ICS information into the reconstructed image using data acquired with the AX-PET prototype. In almost all cases an increase in image quality was observed, particularly for larger features. A small resolution penalty was also observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.