Abstract
The Prolene Hernia System (PHS) is already widely in use in the United Kingdom for inguinal hernias. We describe the novel technique of using the three-in-one design of the PHS (Ethicon Endo-Surgery, Bracknell, UK) for repairing incisional, epigastric and umbilical herniae. This is a three-dimensional device and consists of an onlay patch, a tubular connector and an underlay patch. We recommend a four 'corner' suturing of the underlay patch under vision (and then) through the full thickness of abdominal wall layers to ensure a flat underlay mesh. These four sutures flatten out the underlay patch and can be tied or removed with equal effect. The sutures are placed at 3, 6, 9 and 12 o'clock, which simplifies the procedure and ensures that the underlay lays correctly and is corrugation-free and tension-free, thereby providing a two-layer repair for those herniae with a high rate of recurrence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.