Abstract
BackgroundAn airway-associated fire in an operating room can have devastating consequences for patients. Breathing circuit warmers (BCWs) are widely used to provide heated and humidified anesthetic gases and eventually prevent hypothermia during general anesthesia. Herein, we describe a case of a BCW-related airway fire.Case presentationIn this case, an electrical short within a BCW wire caused a fire inside the circuit. Simultaneously, the fire was extinguished, ventilation was stopped, and the endotracheal tube was disconnected from the BCW. The patient was exposed to the fire for less than 10 s, resulting in burns to the trachea and bronchi. Immediately after airway burn, bronchoscopy showed no edema or narrowing except for soot in the trachea and both main bronchus. After the inhalation burn event, prophylactic antibiotics, bronchodilator, mucolytics nebulizer, and corticosteroid nebulizer were started. On bronchoscopy 3 days after the inhalation burn, mucosal erythematous edema was observed and the inflammatory reaction worsened. The inflammatory reaction showed aggravation for up to 2 weeks, and then gradually recovered, and the epithelium and mucous membrane of the upper respiratory tract returned to normal after 4 weeks. Eventually, the patient recovered without long-term complications and was successfully discharged.ConclusionsThis is the first report of a fire caused by BCW. We wanted to share our experience of how we responded to an airway-related fire in an OR and treated the patient. It cannot be overemphasized that the electrical medical appliance associated with the airways are fatal to the patient in the event of a fire, so caution should always be exercised.
Highlights
An airway-associated fire in an operating room can have devastating consequences for patients
This is the first report of a fire caused by Breathing circuit warmer (BCW)
We wanted to share our experience of how we responded to an airway-related fire in an Operating room (OR) and treated the patient
Summary
This is the first report of a fire caused by BCW. We wanted to share our experience of how we responded to an airway-related fire in an OR and treated the patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.