Abstract

We are performing clinical studies on breast cancer examinations at Hokkaido University Hospital with an ultrasound computed tomography (USCT) system. Our studies have revealed that some reflection images exhibit intensity inhomogeneity because ultrasound waves, shot by a 1-D ring array transducer, go non-vertically into the object surface. This trend significantly increases the burden of interpretation. Therefore, we developed a calibration method to remove this heterogeneity based on the distribution of the incident angle of waves that are estimated from the slope of the subject surface morphologically extracted from multi-slice reflection images. Results showed that applying this correction method to clinical images enabled the image contrast and uniformity to be successfully recovered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.