Abstract

This report provides retrospective follow-up data on 324 consecutive patients who received a Hancock-I porcine valve in the aortic or the mitral position, or in both positions, between June, 1974, and December, 1976. This analysis included 319 valves (193 mitral, 126 aortic) available for study of the incidence of primary tissue valve failure after 10 to 12.5 years of follow-up. Of the 319 prostheses at risk, 114 instances of primary tissue valve failure occurred. Seventy-three of the failed valves were in the mitral position, and 41 were in the aortic position. The calculated actuarial probability of freedom from primary tissue valve failure was 52 ± 5% for the mitral and 58 ± 6% for the aortic prostheses at 12.5 years of follow-up. For patients older than 40 years at the time of operation, the rate of freedom from primary failure was 68 ± 8% and 55 ± 6% for aortic and mitral prostheses, respectively, at 12.5 years. Comparison of both actuarial curves disclosed no meaningful difference. However, a tendency toward greater failure rate was observed in the mitral prosthesis group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.