To evaluate the incidence of intraoperative complications using the NIDEK MK-2000 microkeratome during flap creation in LASIK. The incidence of intraoperative flap complications during LASIK using the NIDEK MK-2000 microkeratome was retrospectively studied in 26,600 procedures performed in 5 outpatient excimer laser surgery centers. A total of 65 (0.244%) complications were identified: 23 (0.086%) eyes had free caps, 13 (0.049%) eyes had an incomplete pass, 13 (0.049%) eyes had an epithelial defect, 11 (0.041%) eyes had buttonhole, and 5 (0.019%) eyes had irregular flap. The remaining 26,535 (99.756%) eyes had uneventful flap creation. Use of the NIDEK MK-2000 microkeratome resulted in a low incidence of intraoperative complications, making it a safe and reliable device for creating flaps during LASIK.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.