To report the incidence of opacification in a consecutive series of 10 eyes that underwent scleral fixated Akreos AO60 placement with concurrent or subsequent contact with gas or silicone oil. Consecutive case series. Three cases of IOL opacification were noted. Two cases of opacification occurred in patients that underwent subsequent retinal detachment repair with C3F8 and one with silicone oil. One patient underwent explanation of the lens due to visually significant opacification. Scleral fixation of the Akreos AO60 IOL is associated with risk of IOL opacification when exposed to intraocular tamponade. While surgeons should consider the risk of opacification in patients at high risk of requiring intraocular tamponade, only 1 in 10 patients developed IOL opacification significant enough to require explantation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.