Abstract

The subcutaneous implantable cardioverter-defibrillator (S-ICD) is a novel technology with proven efficacy in sudden cardiac death prevention; however, there is a lack of long-term safety data. We describe the case of a 55-year-old female patient implanted with an S-ICD due to idiopathic ventricular fibrillation, who subsequently presented with inappropriate shocks leading to ventricular fibrillation that was successfully terminated by another shock. Inappropriate shocks were due to intermittent T wave oversensing during periods of rate-dependent right bundle branch block. Assessment of the S-ICD electrograms during an exercise test allowed successful reprogramming of the device׳s sensing vector with no further events.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.