Abstract

The accuracy of pulse oximetry was studied in a group of patients with severe tricuspid regurgitation. Measurements of arterial oxygen saturation from a finger and an ear probe were compared with those from a radial arterial blood sample analysed in vitro. Lower values were obtained using the pulse oximeter; the difference ranged from +2% to -11%. The discrepancies between pulse oximeter and laboratory oximeter readings were greater in this group of patients than in a control group who did not have tricuspid regurgitation. There was, however, no correlation between the magnitude of this discrepancy and either the peak central venous pressure or the venous pulse pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.