Abstract

Deflated balloon profile is an important consideration in the selection of PTCA catheters. In order to test the accuracy of manufacturer-reported balloon profiles, we measured deflated shoulder and mid-balloon profile in 107 unused catheters (2.0, 2.5, and 3.0 mm) from 6 manufacturers using a precision hole gauge with a resolution of .001 inches. A significant difference was defined as a discrepancy greater than or equal to 0.003 inches from the manufacturer's reported profile. During Phase I, measurements were obtained at room temperature on catheters directly out of the package. Of the 36 models tested, 21/36 (56%) had measured shoulder profiles that were larger than their reported profile, and 29/36 (81%) had mid-balloon profiles that were larger than their reported profile. During Phase II, measurements were made at body temperature following a 60 second inflation at 120 psi (8.16 atm). Of 33 models tested, 18/33 (55%) had measured shoulder profiles that were larger than their reported profile; 26/33 (79%) had mid-balloon profiles that were larger than their reported profile. The term "profile" is imprecisely defined. Manufacturer-reported deflated PTCA balloon profiles are not always accurate. Independent and uniform testing of balloon profiles is necessary for valid comparisons between balloons and manufacturers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.