Abstract

We present a complication associated with herniation of a specially designed silicone-based endotracheal tube (ETT) cuff used for laryngeal nerve monitoring (electromyographic [EMG] ETT). Lung ventilation in our patient was initially unimpeded. However, 30 minutes after initiation of anesthesia with desflurane and nitrous oxide, there was a sudden inability to ventilate due to the herniation of the ETT cuff presumably caused by diffusion of nitrous oxide into the silicone-based cuff. In vitro testing has shown that the increase in intracuff pressure during ventilation with nitrous oxide in the silicone-based EMG ETT is much greater (approximately 50 mmHg) than that in the polyvinyl chloride–based ETT (approximately 10 mmHg) routinely used in our department.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.