Abstract

We performed a retrospective study to illustrate the challenges with quantifying monoclonal (M)-protein in the cases of serum protein capillary zone electrophoresis (SPCZE) where no discernable peak is apparent. We retrospectively reviewed 160 serum immunofixation electrophoresis (SIFE) that were performed at Memorial Hermann Hospital-Texas Medical Center between October 2013 and November 2013 and we identified the positive SIFE results. The corresponding SPCZE of the positive SIFE were retrieved and evaluated for the ability to quantify M-proteins in them. We define the ability to quantify M-protein as the ability for the operator of the SPCZE to identify a discernable peak and to be able to manually gate the area under the peak. Twenty-two cases of SIFE detected a monoclonal immunoglobulin. Of the corresponding 22 SPCZE, we could not quantify the M-protein in 6 (27.3%) of the cases. We have shown several cases where we were not able to quantify the M-protein with SPCZE. This poses a challenge in the diagnosis and management of these patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.