Abstract

The fluorine contents of plastics, ranging from about 20 μg·g−1 to 66%, may be measured instrumentally using a conventional research nuclear reactor, an automated sample irradiation and counting system, and a set of well-calibrated, in-house, fluorine standards. Plastics with low to medium fluorine contents may be analyzed using 20F by placing the gamma-ray detector at appropriate distances from the irradiated sample. For high-F plastics, samples may be irradiated in a cadmium lined irradiation site, using 19O and 20F. Counting statistics of <3% translate into reproducibility of measurements within ±3% and analytical accuracies of ±1% to ±10%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.