Abstract

In 50 cemented hip arthroplasties, wear and migration of the polyethylene (PE) cups were measured with radiostereometric analysis for a period of 2 years. Twenty had a normal gamma-in-air–sterilized PE, another 20 had a PE sterilized with 30 000 Gy followed by heat stabilization (Duration; Stryker Orthopaedics, Mahwah, NJ), and 10 had highly cross-linked PE cups irradiated with 100 000 Gy (Crossfire; Stryker Orthopaedics). In the initial 2 months, head penetration (creep) was 63 μm on average for the 3 groups. From 2 to 24 months, the mean proximal head penetration (wear) was 156 μm for standard PE, 138 μm for stabilized PE ( P = .45), and 23 μm for highly cross-linked PE ( P < .001; analysis of variance). The low in vivo wear rate for highly cross-linked cups was not at the expense of higher migration or less favorable clinical outcome and looks promising.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.