Abstract
Optical coherence tomography (OCT) is limited as an intravascular imaging tool because of interference with blood. This study tested a new balloon occlusion-flushing catheter for OCT scanning of stented coronary arteries and compared stent measurements between OCT and intravascular ultrasound (IVUS). Motorized pullback with OCT and IVUS was examined in coronary stents deployed in swine. Quantitative measurements were obtained and compared between both groups. In addition, stent strut thickness was compared among OCT, IVUS and actual measurement. The occlusion catheter successfully provided motorized pullback OCT images in the stented coronary arteries without any complications. There were no differences in calculated lumen volume. However, stent volumes were significantly smaller with OCT than with IVUS ( p < 0.05). OCT significantly underestimated the stent strut thickness compared with the actual measurement. Although OCT underestimates the stent strut thickness, motorized pullback OCT imaging with the occlusion catheter can provide appropriate in-stent images in the porcine coronary arteries. (E-mail: ykawase@partners.org)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.