Abstract

Conventional Doppler methods for blood velocity estimation only estimate the velocity component along the ultrasound beam direction. This implies that a Doppler angle under examination close to 90 degrees results in unreliable information about the true blood direction and blood velocity. The novel method transverse oscillation (TO), which combines estimates of the axial and the transverse velocity components in the scan plane, makes it possible to estimate the vector velocity of the blood regardless of the Doppler angle. The present study evaluates the TO method with magnetic resonance phase contrast angiography (MRA) by comparing in vivo measurements of stroke volume. Eleven healthy volunteers were included in this prospective study. From the obtained data sets recorded with the 2 modalities, vector velocity sequences were constructed and stroke volume calculated. Angle of insonation was approximately 90 degrees for TO measurements. The correlation between the stroke volume estimated by TO and MRA was 0.91 (p < 0.01) with the equation for the line of regression: MRA = 1.1.TO-0.4. A Bland-Altman plot was additionally constructed where the mean difference was 0.2 ml with limits of agreement at -1.4 ml and 1.9 ml. The results indicate that reliable vector velocity estimates can be obtained in vivo using the presented angle-independent 2-D vector velocity method. The TO method can be a useful alternative to conventional Doppler systems by avoiding the angle artifact, thus giving quantitative velocity information.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call