Abstract

The counting efficiency and detection limit of an in vivo thyroid 131I monitoring system using an imaging plate (IP) were estimated using an anthropomorphic thyroid neck phantom. The counting efficiency of the IP system was approximately constant for thyroid volumes between 11.7 and 20 ml and neck diameters between 10 and 14 cm. The detection limits were distributed from 288 to 451 Bq depending on the combination of neck diameter, thyroid volume and tissue thickness. The IP monitoring system gives a reliable counting efficiency notwithstanding the variation of thyroid volume and neck diameter. The IP system is a new option for thyroid 131I monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.