Abstract

In vivo quantification of radiopharmaceuticals has great potential as a tool in developing new drugs. We investigated the accuracy of in vivo quantification with multi-pinhole single-photon emission computed tomography (SPECT) in rats. Fifteen male Lewis rats with different stages of renal dysfunction were injected with 50 MBq 99mTc-dimercaptosuccinic acid. Four to six hours after injection, SPECT of the kidneys was acquired with a new four-headed multi-pinhole collimator camera. Immediately after imaging the rats were sacrificed and the kidneys were counted in a gamma-counter to determine the absorbed activity. SPECT data were reconstructed iteratively and regions of interest (ROIs) were drawn manually. The absolute activity in the ROIs was determined. Uptake values ranging from 0.71% to 21.87% of the injected activity were measured. A very strong linear correlation was found between the determined activity in vivo and ex vivo (r2=0.946; slope m=1.059). Quantification in vivo using this multi-pinhole SPECT system is highly accurate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.