Abstract

The purpose of this study was to demonstrate the feasibility of computing three-dimensional relaxation maps of spin-lattice relaxation time in the rotating frame (T1rho) from in vivo magnetic resonance (MR) images of the human patellofemoral joint. T1rho was measured by applying a three-dimensional gradient-echo pulse sequence in six healthy subjects and one symptomatic subject by using a 1.5-T MR imager and a 15-cm-diameter transmit-receive quadrature birdcage radiofrequency coil. Average T1rho measured in healthy patellar cartilage was 49.7 msec +/- 3.2 (mean +/- SD). Two-dimensional T1rho-weighted images were obtained with a fast spin-echo pulse sequence for comparison. There was good correlation between two-dimensional and three-dimensional T1rho values for the six healthy subjects (R2 = 0.88, slope = 1.16).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.