Abstract

To evaluate the precision of the iDXA for total body composition and fat distribution measurements in severely obese patients. Sixty-five severely obese participants with a mean age of 46 ± 11 years, BMI of 49 ± 6 kg/m(2) , and a mean body mass of 137.3 ± 20.9 kg took part in this investigation. Two consecutive iDXA scans with repositioning of the total body were conducted for each participant. The coefficient of variation (CV), the root-mean-square (RMS) averages of standard deviations of repeated measurements, the corresponding 95% least significant change, and Intraclass Correlations (ICC) were calculated. Precision expressed as % CV, for total body bone mineral content, fat free mass, total body fat, total body lean, and % total body fat were 1.08%, 0.94%, 0.90%, 1.00%, 0.79%, respectively. Precision was 1.44% for gynoid fat distribution and 1.64% for android fat (AF) distribution. The ICCs in all DXA measurements were 0.99 with % AF having the lowest at 0.96. The GE Lunar iDXA™ demonstrated excellent precision for total body composition assessments and is the first study to assess reproducibility in severely obese individuals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.