Abstract
To evaluate the precision of the iDXA for total body composition and fat distribution measurements in severely obese patients. Sixty-five severely obese participants with a mean age of 46 ± 11 years, BMI of 49 ± 6 kg/m(2) , and a mean body mass of 137.3 ± 20.9 kg took part in this investigation. Two consecutive iDXA scans with repositioning of the total body were conducted for each participant. The coefficient of variation (CV), the root-mean-square (RMS) averages of standard deviations of repeated measurements, the corresponding 95% least significant change, and Intraclass Correlations (ICC) were calculated. Precision expressed as % CV, for total body bone mineral content, fat free mass, total body fat, total body lean, and % total body fat were 1.08%, 0.94%, 0.90%, 1.00%, 0.79%, respectively. Precision was 1.44% for gynoid fat distribution and 1.64% for android fat (AF) distribution. The ICCs in all DXA measurements were 0.99 with % AF having the lowest at 0.96. The GE Lunar iDXA™ demonstrated excellent precision for total body composition assessments and is the first study to assess reproducibility in severely obese individuals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.