To evaluate the thermal load of ablation in high-speed laser corneal refractive surgery with the AMARIS excimer laser (SCHWIND eye-tech-solutions). Thermal load from refractive corrections on human corneas using a 500-Hz laser system with a fluence of 500 mJ/cm(2) and aspheric ablation profiles was recorded with an infrared thermography camera. Each single in vivo measurement was analyzed and temperature values were evaluated. Overall, the maximum temperature change of the ocular surface induced by the refractive ablations was ≤4°C. The increase in the peak temperature of the ocular surface never exceeded 35°C in any case. This low thermal load was independent of the amount of correction the eye achieved. The thermal load of the ablation in high-speed laser corneal refractive surgery was minimized using a computer algorithm to control the peak temperature to avoid corneal collagen denaturation with minimal compromise on treatment duration.