Abstract

A new Roentgen Stereophotogrammetric Analysis (RSA) system is reported; it can measure penetration of the metallic femoral component of a Total Knee Replacement (TKR) prosthesis into the polyethylene bearing on the tibial component. This system was used to analyse a study group of six Anatomic Graduated Components (AGC) knee prostheses more than 6 years post-implantation, and to compare with a control group of six newly implanted AGC prostheses. The volumetric loss of polyethylene was estimated by imaging each prosthesis at a series of different knee flexion angles. The mean difference between the RSA measured polyethylene bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (SD 0.17).The estimated linear penetration at 6.4 years in this prosthesis was determined to be 0.1 mm/year. Volumetric wear was estimated to be 600–700 mm3/year at 6.4 years, equating to ∼100 mm3/year. This does not appear to be clinically significant amount of wear as this prosthesis has excellent survival at 10 years.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.