Abstract
The factors affecting the accuracy and minimum detectable concentration of in vivo tibia lead measurement are discussed, and it is demonstrated that the use of a 109Cd source in a backscatter geometry and using the 88 keV coherently scattered photon for normalisation optimises both criteria. The measurement is shown to be independent of variations in source-sample distance, thickness of overlying tissue and tibia size and shape. Applying the same technique in vitro to samples of human tibia and metatarsals, it is shown that the results are not significantly different (p approximately=0.9) from atomic absorption spectrometry results from another laboratory. The results of Monte Carlo dose distribution calculations are presented and compared with measurements using thermoluminescent dosemeters: the mean absorbed dose to a 20 cm leg section is less than 0.1 mGy (10 mrad) and the maximum absorbed skin dose is 0.45 mGy (45 mrad). For this dose the minimum detectable lead concentration is 10 mu g g-1. Finally, the technique has been applied to groups of normals and occupationally exposed workers, and the means have been shown to be significantly different, namely 10 and 31 mu g g-1 respectively. In the normal subjects tibia lead correlated strongly with age (r=0.63, p<0.001).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.