Abstract

A method has been described and validated for obtaining a sample of blood, without separate venapuncture, which has the same urea and creatinine composition as systemic venous blood during hemodialysis. Using this technique, recirculation in vivo during "Y" type single needle dialysis, measured in 20 bovine fistulas, ranged from 6.9% to 56.5% and averaged 19.4% of blood flow rate as calculated from blood urea nitrogen; and averaged 22.0% with a range of 3.3% to 62.1% of blood flow rate as calculated from serum creatinine concentration. These results suggest that fistula puncture methods and devices that eliminate recirculation in the ex vivo blood circuit should be used to maximize dialysis efficiency.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.