Abstract

Complete catheter-tissue contact and permanent tissue destruction are essential for efficient radio-frequency ablation (RFA) during cardiac arrhythmia treatment. Current methods of monitoring lesion formation are indirect and unreliable. The purpose of this study is to evaluate the feasibility of using optical coherence tomography (OCT) catheter to image endocardial wall in actively beating hearts through percutaneous access. We reported the first in vivo intracardiac OCT imaging through percutaneous access with a thin and flexible OCT catheter. This is a critical step toward image-guided RFA in a clinical setting. A cone-scanning forward-viewing OCT catheter was advanced into beating hearts through percutaneous access in four swine. The OCT catheter was steered by an introducer to touch the endocardial wall. We are able to acquire high quality OCT images in beating hearts, observe the polarization-related artifacts induced by the birefringence of myocardium, and readily evaluate catheter-tissue contact. The observations indicate that OCT could be a promising technique for in vivo guidance of RFA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.