Abstract

This paper describes the results obtained from implanting a strain gauged version of an ‘English’ hip joint replacement together with a totally implantable FM radio transmitter. The implant is based upon a new concept in the design of femoral hip components having a diminished head offset to reduce head load and improved stem shape permitting alignment of the neck along the theoretical axis of peak load transmitted during the gait cycle. The implant was inserted using the ‘English’ trochanteric approach (English, 1975) which further reduces the load on a prosthetic hip joint with the use of a spacer out from the redundant femoral head to rearrange the trochanteric muscle lever arms. The resulting axial load is detected by four strain gauges mounted on a ‘piston in cylinder’ arrangement contained within the thickened neck of the prosthesis. The single channel FM transmitter relays the gauge output to a signal processing unit to give a direct output of activity for recording on a UV recorder. Recordings were taken during implantation, recovery, walking (at three days) physiotherapy, stair climbing and walking over a period of forty days.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.