Abstract
Two anthropomorphic phantom breasts and six patients with breast carcinoma were irradiated according the prescriptions of the EORTC protocol 22881 on the conservative management of breast carcinoma by tumorectomy and radiotherapy. During the implantation procedure for an iridium-192 boost, three tubes were implanted, enabling the measurement with TLD rods of the dose within the breasts of the phantom and the patients during one fraction of the external x-ray therapy and during the interstitial therapy. Measured doses were compared with calculated values from a 2-D dose planning system. In general a fair agreement was found between the measured and calculated doses in points within the breast for the external beam therapy as well as for the interstitial treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.