Abstract

To evaluate stent lumen assessment of various commonly used and newly developed stents for the superficial femoral artery (SFA) using MR angiography (MRA) at 1.5 and 3 T. Eleven nitinol stents and one cobalt-chromium stent were compared regarding stent lumen visualisation using a common three-dimensional MRA sequence. Maximum visible stent lumen width and contrast ratio were analysed in three representative slices for each stent type. A scoring system for lumen visualisation was applied. Nitinol stents showed significantly better performance than the cobalt chromium stent (P < 0.05) at 1.5 and 3 T. Maximum visible stent lumen ranged between 43.4 and 95.5 %, contrast ratio between 7.2 and 110.6 %. Regarding both field strengths, seven of the nitinol stents were classified as "suitable". Three nitinol stents were "limited", and one nitinol stent and the cobalt chromium stent were "not suitable". Intraluminal loss of signal and artefacts of most of the SFA stents do not markedly limit assessment of stent lumen by MRA at 1.5 and 3 T. MRA can thus be considered a valid technique for detection of relevant in-stent restenosis. Applied field strength does not strongly influence stent lumen assessment in general, but proper choice of field strength might be helpful.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.