Abstract
The image quality of dynamic 3-D echocardiography is limited by temporal jitter artifacts that result from the asynchronous acquisition of video frames with the cardiac cycle. This paper analyzes the source and extent of these artifacts using in vitro studies. Dynamic 3-D images of a myocardial motion phantom were reconstructed and analyzed for eight cardiac motion patterns. The extent of temporal jitter artifacts was quantified, first, from the images by computing temporal jitter maps and, second, predicted from the motion waveforms. Temporal jitter appeared as a pattern of streak artifacts converging at the axis of rotation of the imaging plane, for the rotational scanning approach used in our study. The results of the experimental analysis techniques were compared with the waveform analysis using linear regression analysis. The least squares line showed good correlation between the data (r > 0.9) and its deviation from the line of identity was calculated to be <9%. (E-mail: afenster@irus.rri.on.ca)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.