Abstract

We specially devised 3 types of turbo pumps, a centrifugal pump (CFP), a mixed flow pump (MFP), and an axial flow pump (AFP), and analyzed their in vitro performance. The common structural design elements were an impeller diameter of 20 mm and sealless magnet couple driving. In vitro tests were carried out using heparinized fresh bovine blood. The hemolysis was comprehensively evaluated at 7-16 points by changing the flow rate and pressure head (mapping of hemolytic property). The maximum efficiency (motor output to pump output) was 44.9% at 7,000 rpm, 3.17 L/min, 191 mm Hg in the CFP; 66.3% at 7,000 rpm, 6.9 L/min, 136 mm Hg in the MFP; and 20.6% at 9,000 rpm, 5.54 L/min, 74 mm Hg in the AFP, respectively. The minimum normalized index of hemolysis (NIH) (g/100 L) was 0.038 at 5,000 rpm, 4.60 L/min, 38 mm Hg in the CFP; 0.010 at 7,000 rpm, 8.22 L/min, 100 mm Hg in the MFP; and 0.033 at 7,000 rpm, 2.84 L/min, 48 mm Hg in the AFP, respectively. The best efficiency and NIH were achieved in the MFP.