Using an organotypic culture method, we evaluated the biocompatibility of two kinds of intraocular lenses: conventional poly(methyl methacrylate) (PMMA) and heparin-surface-modified (HSM) PMMA. Chicken corneal endothelium from embryos incubated for 14 days was placed on an agar medium and covered with one of three types of materials: PMMA, HSM PMMA, and a control, Thermanoxa. Experiments at five different times measured cell migration, cell multiplication, and cell adhesion. Scanning electron microscopy showed a low level of corneal endothelial adhesion on the HSM IOL surface. There was a significant difference between the HSM and untreated lenses in migration surface and cell density, with the HSM lenses having better biocompatibility.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.