Abstract

Purpose: Turbulent flow just distal to stenoses causes signal loss (dephasing) on magnetic resonance angiography (MRA). This study correlates dephasing with trans-stenotic pressure gradients in an in vitro model of arterial stenosis. Materials and methods: Three-dimensional (3D) phase contrast, 2D time-of-flight, and 3D spoiled gradient echo MRA with/without gadolinium and varied echo time were performed for a system consisting of a peristaltic perfusion pump and a silastic vessel with stenoses of varying caliber. Length and diameter of dephasing jets were measured, and volumes calculated at varying pressure gradients and echo times, then correlated with percentage cross-sectional area stenosis as measured by conventional angiography. Results: Dephasing occurred in all sequences at pressure gradients of ≥4 mmHg (1 mmHg = 133 Pa) and stenoses of greater than 70%, and varied directly with pressure gradient. The dephasing was greatest for 3D phase contrast (PC). Gadolinium did not diminish dephasing. Conclusions: MRA signal dephasing at stenoses varies directly with pressure gradient. MRA may provide a non-invasive means for determining the hemodynamic significance of arterial stenoses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.