Abstract
Respimat soft mist inhaler (SMI) is an inhaler producing a soft mist with low velocity. The liquid drug solution is emitted by the mechanical energy of compressed spring, so the device is propellant-free. This paper describes the measurement of particle fraction penetrating through the model of the upper respiratory tract (mouth cavity, pharynx, larynx trachea). The solution inside the cartridge was marked by fluorescein in order to evaluate the regional particle deposition in the model. After the exposition, segments of the model, as well as the filter, were washed out and the deposited fluorescein was extracted into water. The analysis showed that the fraction of particles penetrating through the upper respiratory tract was 64.7±6.7 %. However, the real inhalation pattern was not simulated and measurement was provided with continual “inhalation” without “exhalation”.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.