Abstract

To test in vitro the feasibility of welding amniotic membranes using Nd:YAG laser energy. Fresh fetal membranes from term pregnancies were washed and cut into 1 cm2 pieces. Pooled cryoprecipitate (CPT), 50% albumin (Alb), or polytetrafluoroethilene (e-PTFE) were used as solder medium. The optimal settings of the laser were determined. Results were assessed quantitatively and semi-quantitatively using Pearson Chi-square analysis. Laser welding of amniotic membranes was successful in 82.6% of experiments with e-PTFE and in 10.7% of experiments with CPT (P < 0.001). The strength of the welding was also significantly better with e-PTFE (P < 0.001). Optimal results were obtained using 1-7 Watts and 0.1-1 seconds. Laser welding was unsuccessful in 100% of experiments with Alb. Laser welding of fetal membranes can be accomplished with e-PTFE and to a lesser degree with the CPT using Nd:YAG energy under low wattage-high exposure settings. Further studies are underway to test other grafting or soldering materials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.