Abstract

Objective:The aim of this study was to evaluate temperature alterations on the outer root surface during post space preparation with six different post drills by using an infrared thermometer.Materials and Methods:Sixty extracted single-rooted human mandibular incisor teeth were used. After root canal obturation, the specimens were divided into six groups (n = 10). During post space preparation, the temperature rises were measured in the middle third of the roots using a noncontact infrared thermometer with a sensitivity of 0.1°C. The temperature data were transferred from the thermometer to the computer and were observed graphically.Results:The maximum temperature rise was observed in Snowpost 2 (29.95 ± 10.2°C) (P < 0.001), but there were no significant differences among Snowpost 2 (29.95 ± 10.2°C), Snowpost 1 (24.6 ± 8.0°C), and Relyx 2 (17.68 ± 9.1°C) (P > 0.05).Conclusions:Although water coolant used, the critical temperature rise was observed on the outer root surface in all post drill systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.