Abstract

Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imaging-related heating, and artifacts. Therefore, we evaluated these risks for aneurysm clips. Three different MP35N aneurysm clips (Codman Slim-Line Aneurysm Clip, straight, blade length 25-mm; Codman Slim-Line Aneurysm Clip Graft, 5-mm diameter x 5-mm width; Codman Slim-Line Aneurysm Clip, reinforcing 30 degrees angle, 6-mm x 18-mm) that represented the largest mass for 155 additional clips made from MP35N were tested. The clips were evaluated at 3T for magnetic field interactions, heating, and artifacts. We studied MR imaging-related heating, placing the clip in a gelled-saline-filled phantom with MR imaging performed by using a transmit/receive radio-frequency body coil at a whole-body average SAR of 3 W/kg for 15 minutes. Artifacts were characterized by using T1-SE and GRE pulse sequences. Each aneurysm clip showed relatively minor magnetic field interactions, which would not cause movement in situ. Heating was not excessive (highest temperature change, <1.8 degrees C). Artifacts may create problems if the area of interest is in the same area or close to the aneurysm clip. The results of this investigation demonstrated that it would be acceptable (ie, "MR conditional" using current terminology) for patients with these aneurysm clips to undergo MR imaging at < or =3T. Notably, on the basis of the sizes of the clips that underwent testing, these findings pertain to 155 additional aneurysm clips made from the same material.

Highlights

  • AND PURPOSE: Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imagingϪrelated heating, and artifacts

  • While the presence of a ferromagnetic aneurysm clip in an individual is a strict contraindication for an MR imaging procedure,[3,4,5,6,7,8] aneurysm clips classified as “nonferromagnetic” or “weakly ferromagnetic” cause no known MR imagingϪrelated problems.[5,6,7,8,9,10,11,12,13,14,15,16,17]

  • Biomedical implants pose possible risks and other concerns for patients referred for 3T MR imaging examinations, which include movement or displacement of the device, excessive heating of the object, and substantial artifacts, which may impact the diagnostic use of MR imaging.[7,8]

Read more

Summary

Introduction

AND PURPOSE: Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imagingϪrelated heating, and artifacts. We evaluated these risks for aneurysm clips. Codman Slim-Line Aneurysm Graft Clip, 5-mm diameter ϫ 5-mm width 1 2 3 Background Codman Slim-Line Aneurysm Clip, reinforcing 30°. Ϩ1.7 ϩ1.7 ϩ1.7 ϩ1.6 each clip, with the GRE pulse sequence producing larger artifacts than the T1-SE pulse sequence.

Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.