Abstract
We report on methodologies for use in the design of a biodegradable Mg alloy appropriate for load-bearing but temporary orthopedic implant applications. Comparative studies of Mg-5Ca and Mg-5Ca-1Zn were conducted to explore the effects of a combination of minor alloying and hot extrusion, on the alloy's mechanical properties and corrosion resistance. The extruded Mg-5Ca-1Zn exhibited high ultimate compressive strength of 385 MPa and suffered no significant structural degradation even after immersion in simulated body fluid for 30 days. Mg-5Ca-1Zn alloy showed the mechanical strength and controlled corrosion rate to be considered as an ideal candidate for biodegradable orthopedic implant material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Biomedical Materials Research Part B: Applied Biomaterials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.