Abstract
Experiments were performed to determine whether a clinical diagnostic scanner, a Hewlett-Packard (HP) 77020A, could produce cavitation in water containing suspensions of either 0.245-mum polystyrene spheres or Albunex, 1-10 mum albumin-coated microbubbles. Two calibrated, phased-array HP imaging transducers with 2.5- and 5.0-MHz operating frequencies were driven in M-mode (single cycle) and Doppler mode 4 cycles by the HP imaging system. Cavitation was detected in the water with polystyrene spheres at 2.5 MHz in both M-mode and Doppler mode at a peak negative acoustic pressure of 1.1 MPa or greater. Insonification at 5.0 MHz in either mode did not produce a detectable amount of cavitation, even with peak negative pressures as high as 1.2 MPa. Cavitation was not detected in water with the Albunex spheres at either frequency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.