Abstract

AbstractBiodegradable magnesium alloys are promising materials for orthopaedic implants. With their mechanical properties similar to native bone, magnesium alloys can overcome some significant disadvantages of conventional metallic implants. Nonetheless, the material degrades too fast for clinical approval. Therefore, the corrosion rate of the implant needs to be controlled and decelerated to guarantee sufficient mechanical support during the entire bone healing process. In this work, we analyse the corrosion behaviour of the magnesium alloy WE43 in a physiological environment and assess the influence of a plasma electrolytic coating to the corrosion process. The experimental results show that the coating significantly decreases the mass loss rate as well as the degradation of the material strength of the WE43 alloy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.