Abstract

The purpose of this study was to determine whether or not prefilled breast implants retain their volume in packaging. This study examined 32 Poly Implant Prosthesis prefilled textured saline breast implants. All of these implants were within the manufacturer's expiration date at the time of weighing. No holes were visible on any of the implants. All were weighed on an electronic scale. The measured weight was compared with the expected weight (based on the implant size as specified by the manufacturer) and the percentage deflation was calculated. The manufacturer declined to provide specific information about the manufacture dates of the implants; thus, relative age (rather than absolute age) was examined with respect to percentage deflation. Of the 32 implants examined, all showed some degree of deflation (range, 8.84 to 57.14 percent; 95 percent confidence interval of the mean, 22.01 +/- 4.17 percent). There was a moderate correlation (r = 0.41) between relative age of the implant and percentage deflation. From these results, it is clear that this type of prefilled saline breast implant does not maintain its volume in vitro. If these implants are used, the underfilling could contribute to a higher deflation rate and cosmetic deformity. This risk should be taken into account by plastic surgeons who use this type of implant in breast augmentation procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.