Abstract

Objectives. To investigate the in vitro electrochemical corrosive behavior of archwires used in lingual orthodontics and the effects on the phase transition temperatures.Materials and Methods. Six different types of archwires of stainless steel, titanium-molybdenum, nickel-titanium and nickel-titanium-copper were used. Corrosion tests were performed following ISO-standard 10993-15:2000. Differential scanning calorimetry and scanning electron microscopy were used.Results. The stainless steel archwires showed anEpitaround −600 mV, and those of titanium alloys showedEpitvalues around 1000 mV. Differential scanning calorimetry detected a rhombohedral phase in nickel-titanium archwires, while it was not detected in nickel-titanium-copper wires. A difference of 2°C to 3.5°C from the manufacturer's claim was found in the as-received and polarized samples, respectively.Conclusions. The 0.016 stainless steel archwires were found to be the less resistant to corrosion. A rhombohedral phase was detected on the nickel-titanium archwires. No major differences were observed among groups concerning phase transformation temperatures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.