Abstract
This study aimed to compare in vitro two syringe changeover techniques to determine which was better at minimising variations in norepinephrine (NE) delivery: the manual quick-change or automated technique. NE concentration was measured continuously using a UV spectrophotometer, and infusion flow rate was monitored by an infusion pump tester. Syringe changeovers were made with either of the two techniques studied. Relays induced disturbances in drug delivery. The temporary increase in NE mass flow rate was significantly higher with manual relays than with automated ones. The automated relay offered a better control of the amounts of NE administered than the quick-change technique.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.