The aim of this randomized in vitro study was to compare the time and accuracy of implant-site preparation and implant placement using a trephine drill versus a conventional drilling technique under dynamic navigation. In total, 42 implants were placed in simulation jaw models with the 2 drilling techniques by 2 operators who had previous experience with dynamic navigation. The timing of each implant placement was recorded, and horizontal, vertical, and angulation discrepancies between the planned and placed implants were compared. There was no significant difference in time or accuracy between the trephine and conventional drilling techniques. Implant-site preparation with a single trephine drill using dynamic navigation was as accurate under in vitro experimental conditions as a conventional drilling sequence.