Abstract
IntroductionExperimental evaluation of endometric devices usually relies on visual, subjective detection of the apical constriction to determine the accuracy of measurements. The aim of the present study was to analyze the accuracy of measurements of Raypex 6 and EndoPilot using a novel, objective image-analysis system.MethodsOnehundred and twenty teeth were randomized and allocated to three groups: After coronal flaring, either Raypex 6 or EndoPilot were used to determine the endodontic working length during instrumentation using manual files (RPM and EPM group respectively). In addition, EndoPilot was used for continuous, automatic measurement during rotating instrumentation (EPA group). If the working length had been reached according to endometric results, the files were fixed in place. Tooth and file were then embedded and prepared for analysis. Subsequently, the distance between the tip of the file and the apical constriction (DAC) or the apical foramen (DAF) was calculated using trigonometric analysis and the position of the file relative to AC and AF was analyzed.ResultsBoth inter- and intra-examiner-reliability of the trigonometric analysis were nearly perfect (ICC = 0.999, p<0.001). DAC was not significantly different between groups (p>0.05, t-test). DAF was significantly decreased when EPA had been used compared to EPM (p<0.05, Exact-test). EPA resulted in files being positioned beyond AF significantly more often than the other two methods (p<0.01).ConclusionsAll methods allowed reliable detection of AC. However, EPA significantly increased the risk of overpreparation. Objective, digital assessment based on image analysis was suitable to compare the accuracy of different endometric devices.
Highlights
Experimental evaluation of endometric devices usually relies on visual, subjective detection of the apical constriction to determine the accuracy of measurements
DAF was significantly decreased when EPA had been used compared to EPM (p
Digital assessment based on image analysis was suitable to compare the accuracy of different endometric devices
Summary
The aim of the present study was to analyze the accuracy of measurements of Raypex 6 and EndoPilot using a novel, objective image-analysis system. The aim of this study was to compare the accuracy of manual and continuous endometric measurements in the assessment of the endodontic working length using an objective digital analysis system
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.