Abstract

Pediatric patients receiving respiratory support with heated flow nasal cannula (HFNC) systems frequently receive inhaled medications. Most available data have been obtained with vibrating mesh nebulizers that are expensive. Data are lacking regarding the feasibility of using less expensive devices such as continuous output jet nebulizers. The characteristics of the aerosols generated by jet nebulizers operated at different conditions (6 and 9 L/min) were studied alone and connected to a HFNC system and different size cannulas using a cascade impactor and spectrophotometry (276 nm). Aerosol characteristics changed while traveling through the HFNC system. Initial size selection occurred at the exit of the circuit (before connecting to the cannula) with all aerosol <5 µm. Nasal cannula size further selected aerosols and reduced drug delivery. The operating flow of the nebulizer did not affect the delivered mass but higher flows generated smaller particle size aerosols. The addition of supplemental flow significantly reduced the delivered mass. The measured aerosol characteristics would likely result in intrapulmonary deposition. The delivery of aerosolized albuterol generated by a continuous output nebulizer placed in the inlet of a HFNC system and connected to large or XXL cannulas is feasible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.