Abstract

The surface of a silicone-disc intraocular lens (IOL) was hydrophilized by plasma etching (oxygen plasma) and compared to an untreated but otherwise identical IOL. Various methods of surface analysis were used to characterize the modification (e.g., X-ray photoelectron spectroscopy, contact angle estimation). A cytotoxic effect of the modified surface was excluded by cell culture experiments evaluating cell spreading, cell morphology, DNA and protein synthesis. In vivo experiments on rabbits indicated that the postoperative foreign-body reaction was not significantly affected by the hydrophilization of the IOL surface. Throughout the entire followup (12 weeks) we found less induced posterior synechias in the eves with bydrophilized lenses than in those with untreated lenses (P = .009). While the IOL dislocations out of the capsular bag and the posterior capsular opacification rate did not differ significantly between the two groups of eyes, we did see special patterns of posterior capsular opacification on the posterior capsules of eyes with the hydrophilized IOL.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.