Abstract

In this work, rapamycin-eluting poly (d, l-lactic acid) coating (PDLLA/RAPA) was prepared on biodegradable Mg-Nd-Zn-Zr alloy (JDBM) for both in vitro and in vivo investigation of the degradation behaviors of the magnesium alloy stent platform. Electrochemical tests and hydrogen evolution test demonstrated significant in vitro protection of the polymeric coating against magnesium degradation both in short and long term. The 3-month in vivo study on the RAPA-eluting JDBM stent implanted into porcine coronary arteries confirmed its favorable safety, and in the meanwhile revealed similar neointima proliferation compared to the second generation DES Firebird 2 with no occurrence of adverse complications. Moreover, Micro-CT examination combined with IVUS and OCT detection indicated a remarkably lower degradation rate and prolonged radial supporting duration of the drug-eluting JDBM stent as compared to the bare, attributable to the protection of the coating in vivo. Hence, rapamycin-eluting JDBM stents exhibit great potential for clinical application.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.